Description
- For the technical ventilation of safety cabinet type 90 in accordance with DIN EN 14470-1
- Reacts to coarse particulate contamination such as dust and heavy pollution of the exhaust air with hydrocarbon vapours and gases
- Warnings displayed by LED when saturation occurs
- Cabinet exhaust air is cleaned of hydrocarbon vapours by integrated active carbon filters and returned to the room air
- Filter is constructed in a tandem design. If the main filter is saturated, the second integrated filter ensures additional safety over a time period of around 14 days.
- The use of a GFA eliminates the compliance with an explosive zone around the safety cabinet
- Exhaust air line to the outdoors and the associated installation work can be omitted
- No pollution of the room air thanks filtering of almost 100% of all solvent fumes
- Increased employee health protection
- Low servicing and maintenance costs
Note:
- To ensure that the filter system is operating correctly, the GFA should be inspected for correct function and serviced at least once a year by a qualified person.
- The tandem active carbon filter must be replaced as part of the inspection
Housing: | Powder-coated sheet steel | |||||||||
Colour: | Light grey RAL 7035 | |||||||||
Dimensions(W× D × H): | 310 × 510 × 210 mm | |||||||||
Motor power: | 0.03 kW | |||||||||
Rated voltage: | 230 V | |||||||||
Frequency: | 50 Hz | |||||||||
Current consumption: | 0.2 A | |||||||||
Noise level: | <> | |||||||||
Empty weight: | 13 kg | |||||||||
Volumetric flow: | 25 m²/h |
90719012001
K_18_115_1